Model 6517B Electrometer Reference Manual

:CNDisplay

This command clears (cancels) NEXT display.
Usage
:DISPlay:CNDisplay
Details
This action command is used to clear the bottom display of NEXT (or PREV) messages and cancel
the operations associated with them. This command has no effect on any other message types.
Also see
None
:SMESsage <b>
This command is used to control the status messages.
Usage
:DISPlay:SMESsage <b>
:SMESsage?
<b>
Details
This command is used to enable/disable the status message mode. The status message mode is a diagnostic tool
that provides real-time messages that relate to the current operating state of the instrument.
Also see
Status and error messages
6517B-901-01 Rev. C / August 2015
0 or OFF
1 or ON
(on page 2-11)
Section 14: SCPI command reference