8.3 Specifications - Test Instructions

8.3.1

Fluorescence Top

Sensitivity
2008-07
®
For the
with the option 'Fluorescence Top', the following tests can
infinite
200
be performed to prove the specifications:
Sensitivity
Uniformity
Precision
These test instructions are valid for the
®
infinite
F200
®
infinite
M200
Standard version
Spectrally enhanced version
Perform the following measurement to determine the detection limit for
Fluorescein:
Measurement Parameters:
Parameter
Reading Mode
Exc Wavelength
Em Wavelength
Number of flashes
Integration Time
Settle Time
Gain
Plate Type
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
8. Quality Control
®
:
infinite
200
Setting
Fluorescence Top
®
: 485 (20 nm)
infinite
F200
®
: 485 nm
infinite
M200
®
: 535 (25 nm)
infinite
F200
®
: 535 nm
infinite
M200
25
40
0
Optimal
GRE96fb