Calibration and Troubleshooting

Step 1: Gain Calibration

Figure 7-2 Calibration: Step 1 Gain Calibration
Adjust the gain so that the real line and ideal line intersect at a
programmed value of 90%.

Step 2: Offset Calibration

Figure 7-3 Calibration: Step 2 Offset Calibration
7-4
Programmed Value
Programmed Value
Adjust the offset so that the real and ideal lines intersect at a
programmed value of 10%.
Programmed Value
Programmed Value
M370430-01 Rev B