Lxi Compliance - Ametek Sorensen XG Operation Manual

Programmable dc power supply 670 watts – 1700 watts
Table of Contents

Advertisement

Communications Options

LXI COMPLIANCE

6-64
The XG is compliant with LXI version 1.2, Class C. For more
information on the LXI standard please visit the LXI Consortium
Home Page. This section describes how certain features of LXI
have been implemented in the XG power supply.
LAN Indicator LED
The PGM LED on the front panel is used as the LAN indicator. If
there is no connection, the LED is off. If there is a connection, the
LED is on. While the unit is attempting to make a connection the
LED blinks. This same LED also blinks when clicking Identify
Device On... on the Web interface to identify which power supply
the user is communicating with.
LAN Reset
In order to return to the factory default settings, a LAN reset may
be performed. To do this
1. Set the 9 position mode control to PGM
2. Turn the rotary encoder knob (left side of front panel display) to
LAN.
3. Press and Hold the rotary encoder knob until the LAN reset
message appears on the front panel 7 segment display
M370430-01 Rev B