Clearing Alarms

Clearing Triggered and Manual Alarms
To clear a triggered alarm, use one of the following methods:
1.
Important: When an alarm is cleared by using this method, only the current
highest precedence alarm is cleared. If another alarm has occurred, then this
alarm will now become the highest precedence alarm and will begin blinking on
the display.
2.
Clearing a Flash Failure Alarm
The Flash Failure alarm is the only exception for clearing alarms. If a
flash failure occurs at boot time, it can only be cleared by a soft reset key
combination press (see "Resetting the Power Supply" on page3-43) or
by power cycling the unit. If a checksum error is detected when recalling
a user slot, no modification of settings will occur, and an error will be
pushed to the SCPI error queue. The alarm will clear after a duration per
the currently configured front panel timeout setting.
To clear a triggered alarm, use one of the following methods:
The SCPI command (s) for these instructions are:
[:]OUTPut[<channel>]:PROTection:CLEar
3-22
Turn the power supply Off and then On.
Press and hold the rotary Adjust/Enter control for 3 seconds.
To clear a manual alarm:
Press and hold the rotary Adjust/Enter control for 3 seconds.
The
message will be displayed on the front panel for a
few seconds.
Repeat the alarm clearing until you have cleared all alarms.
Turn the power supply Off and then On.
For failures that occur during a recall of user memory settings, wait
the duration of the currently configured front panel timeout.
Press and hold the reset key combination (Flash Fail ONLY). See
"Resetting the Power Supply" on page 3-43.
M370430-01 Rev B