Remote Operation

Remote Interface Addressing

All commands must be issued with a multichannel address or the device
must be selected using the:
*adr or :SYST[<channel>]:COMM[:MCH]:ADDR commands.
Once a device is selected all commands sent without a multichannel
address will be handled by the selected device. The use of multichannel
addresses supersedes the selected device as the destination for a
message. For more details on how multichannel addresses affect
responses, see Table 5-8.
To change the selected device simply issue the
[<channel>] :COMM[:MCH] : ADDR
multichannel address. The previously selected device will be deselected
and the new device will be selected. From that point on, all messages
without a multichannel address will be destined for the newly selected
device. It is possible to deselect all devices by using the aforementioned
commands with the broadcast address, 0.
The SCPI Commands for these instructions are:
[:]SYSTem:COMMunicate[:MCHannel]:ADDRess
or
*ADR <address>
Where:
<address> is the multichannel address of the device to be selected. The
address is an integer value from 1 to 30.
5-26
*adr
:SYST
or
command with a different
<address>