Standard Event Status Register (SESR)

The standard event status register sets bits for specific events during
power supply operation. All bits in the standard event status registers are
set through the error event queue. The register is defined by IEEE 488.2
register and is controlled using 488.2 common commands: *ESE,
*ESE?, and *ESR? as well as SCPI aliases for multichannel use.
Standard Event Status Enable Register
The Event Summary Enable command determines which bits in the
Standard Event Status Register are summarized in the Event Summary
Bit (ESB) of the Status Byte. This register will be cleared at power up.
To enable events, you must make the logical sum of all values for the
events that you to be reported and send this parameter in decimal
format.
See Table 5-9 for values for the events to be enabled in <status-enable>
parameter.
For example, sending *ESE 16 sets bit 4 of the Standard Event Status
Enable Register. This will cause the Event Summary bit (ESB) in the
Status Byte to be set whenever the Execution Error bit (bit 4) in the
Standard Event Status Register gets set.
Command:
*ESE
The SCPI equivalent for multichannel use:
[:]STATus[<channel(s)>]:STANdard:ENABle <status-enable>
[:]STATus[<channel(s)>]:STANdard:ENABle?
Standard Event Status Register
Command:
*ESR?
The SCPI equivalent for multichannel use:
[:]STATus[<channel(s)>]:STANdard[:EVENt]?
M370430-01 Rev B
<status-enable>,
The Standard Event Status Register query allows you to determine
the current contents of the Standard Event Status Register. (See
"Standard Event Status Register (SESR)". Reading this register
clears it.
Remote Operation
*ESE?
5
5-35