Remote Operation

Locking and Unlocking the Front Panel

Locking out the front panel will prevent any of the buttons from
functioning. All the buttons and knobs on the front panel will display the
LOCL L
pressed or rotated. This mode prevents any changes to the unit from the
front panel. See the following procedure to lock and unlock the front
panel.
The SCPI Commands for these instructions are:
[:]SYSTem[<channel>]:REMote:STATe {?|LOCal|REMote}

Auto Sequence Programming

5-62
message to be display on the Current and Voltage displays if
OC
Where:
LOCal is used to set the front panel mode to local.
REMote is used to set the front panel into locked mode (remote).
For example:
SYST : REM: STAT
programming mode only).
SYST : REM: STAT
SYST : REM: STAT ?
Auto Sequence programming lets you execute a program stored in
the power supply's memory. The program that can be stored in
memory can be composed of any of the SCPI commands described
in this Manual.
For example, Auto Sequence programming allows you to set the
timer to turn the power output of the supply Off or On after a time
interval or to program voltage sequences.
When a program is activated each command stored in the buffer is
executed as if the operator had sent the command by typing it in. A
variable delay time called dwell is spent between commands. The
dwell time can range from 0 seconds to 180 seconds. See "Setting
Dwell Time" on page 5-63.
The Auto Sequence program will be cleared after AC power is
turned off; however, a procedure exists for storing it in a text file
which can be reloaded at a later date.
REM
Front panel is locked (remote digital
LOC
Front panel is unlocked.
Query for status
M370430-01 Rev B