Replacing The Doors - Thermo Scientific Vanquish VH-D10 Operating Manual

Diode array detectors
Table of Contents

Advertisement

7.12 Replacing the Doors

When
Parts required
Follow these steps
Diode Array Detectors (VH-D10)
Operating Manual
Damage of door
TIP The maintenance procedures do not require that you remove the
doors. If this should ever be required for a specific reason or procedure,
follow the steps in this section.
Replacement door
NOTICE
To avoid damage to the door hinges, be careful when performing the
following sequence of steps and do not apply force.
1. If the door is located directly below the solvent rack, lift the solvent
rack slightly on the front edge.
2. To remove a door, push the door upward while opening.
Open the door to a position in which the two hinges on the housing
are aligned in the grooves on the door. You can remove the door
only when the hinges are in the grooves.
Figure 44: Unhinging a door
No.
Description
1
Hinge on the housing
2
Reception groove on the door
7
Maintenance and Service
Page 151