6.10.3 Restart after Long-Term Shutdown

Diode Array Detectors (VH-D10)
Operating Manual
9. The step depends as follows:
Situation
Device and all other system
modules remain in the
system stack and are to be
turned off
Device shall be transported
or shipped after shutdown
See also
2
Connecting the Inlet Capillary (} page 68)
2
Connecting the Detector Waste Line (} page 69)
To restart the device after a long-term shutdown, follow these steps:
1. Prepare and restart the other modules in the Vanquish system,
following the instructions in the Operating Manuals for the modules.
Pay special attention to the Preparing the Module for Operation
section.
2. Flush the components in the flow path before the flow cell before
you connect the flow cell to the system flow path.
3. Connect the inlet capillary to the flow cell inlet and the waste line to
the flow cell outlet.
4. Turn on the device. Observe the following:
Situation
If the device remained in the
system stack and all system
modules were turned off.
If the device is restarted after
transport.
5. Before starting an analysis, let the detector equilibrate and be sure
that it is ready for operation.
Steps
Turn off the system with the system power button on
the system base.
If one of the modules shall be removed from the
system stack, turn off all system modules with their
main power switch. Pressing the system power
button will not be sufficient to turn off the power to
the devices completely.
Follow the instructions in
Transporting or Shipping
the Device
(} page 153).
Action
Turn on the system with the system power
button on the system base.
Turn on the device with the main power switch.
6
Operation
Page 109