Thermo Scientific Vanquish VH-D10 Operating Manual page 150

Diode array detectors
Table of Contents

Advertisement

7
Maintenance and Service
Page 150
3. Monitor the Audit Trail of the Instrument Configuration Manager
program to see whether the firmware update was successful or
failed.
4. Depends on the situation:
Situation
Firmware update
successful
Firmware update failed
Firmware update fails
repeatedly
Action
Requalification of the device may be required. Refer to
the release notes.
Turn the device off and on again. Repeat the firmware
update.
Contact Thermo Fisher Scientific Technical Support.
Diode Array Detectors (VH-D10)
Operating Manual