7.14.4 Setting Up the Slide-In Module

Follow these steps
Page 160
4. To facilitate tightening the slide-in module in the enclosure, pull out
the module a few millimeters (approximately 2 to 5 mm).
5. With the screwdriver, move each captive screw back and forth while
pressing inward until the screw slips into the thread.
6. Push the slide-in module back into the enclosure as far as it goes in.
7. Gradually and evenly, tighten the four captive screws on the slide-in
module hand-tight.
NOTICE
Verify that the screws are tightened. Pull the slide-in module by the
leak tray towards the front and check whether the screws move. If
they do not move, the slide-in module is installed properly.
If the screws move, tighten the screws further. With a torque
wrench, the recommended torque is 1.2 Nm.
8. Set up and restart the device.
After you have installed the slide-in module in the enclosure, set up and
restart the device.
1. Set up the slide-in module:
a) Connect the slide-in module and set up flow connections (see
Installation
(} page 45)).
b) Prepare the slide-in module for first-time operation (see
Preparing the Device for Operation
c) If you installed a replacement slide-in module, update the
instrument configuration in the chromatography data system
accordingly.
2. Prepare all other modules of the Vanquish system for operation and
restart them. Refer to the Operating Manuals for the modules.
3. Before starting an analysis, let the chromatography system
equilibrate and be sure that it is ready for operation.
4. In the Chromeleon software, run the device-specific Performance
Qualification (PQ) tests.
5. When the test is completed, update the QualificationDone
parameter.
(} page 90)).
Diode Array Detectors (VH-D10)
Operating Manual