Firmware Failure
Diode Array Detectors (VH-D10)
Operating Manual
Messages in the Instrument Audit Trail are preceded by an icon. The icon
identifies the seriousness of the problem. For possible causes and
remedial actions, see
Messages
If a firmware failure occurred during operation of the module, an
exception log has been created about the processes during the firmware
failure. The firmware sends the exception log to the Instrument Audit
Trail when the module is connected in the chromatography data system.
In this case, observe the following:
•
Send the Instrument Audit Trail as .cmbx file to the Technical
Support before you clear the log.
•
To clear the exception log and continue operation of the module,
perform the command ExceptionLogClear.
For more information, refer to the Chromeleon Help.
8
(} page 164).
•
Troubleshooting
Page 163