Diode Array Detectors (VH-D10)
Operating Manual
WARNING—Self-Ignition of Solvents
Solvents with a self-ignition temperature below 150 °C might ignite
when in contact with a hot surface (for example, due to leakage in the
chromatography system).
Avoid the use of these solvents.
WARNING—Hazardous Vapors
Mobile phases and samples might contain volatile or flammable
solvents. The handling of these substances can pose health and safety
risks.
Avoid accumulation of these substances. Make sure that the
installation site is well ventilated.
Avoid open flames and sparks.
Do not operate the device in the presence of flammable gases or
fumes.
CAUTION—Escape of Hazardous Substances from PEEK Capillaries
Some capillaries in the system are made of PEEK. Swelling or attack by
acids can cause PEEK capillaries to start leaking or to burst. Certain
chemicals, for example, trichlormethane (CHCl
(DMSO), or tetrahydrofuran (THF) can cause PEEK to swell. Concentrated
acids, such as sulfuric acid and nitric acid, or a mixture of hexane, ethyl
acetate, and methanol, can attack PEEK.
Swelling or attack is not a problem with brief flushing procedures.
For more information, refer to the technical literature on the
chemical resistance of PEEK.
CAUTION—Allergic Reaction
Some capillaries in the system are made of MP35N™, a nickel/cobalt-
based alloy. Individuals with sensitivity to nickel/cobalt may show an
allergic reaction from skin contact.
2
Safety
), dimethyl sulfoxide
3
Page 25