5.7.5.2
Connecting the Detector Waste Line
Preparations
Follow these steps
Diode Array Detectors (VH-D10)
Operating Manual
2. If the column compartment is located to the right of the detector
Route the capillary through the bottom recess in the partition panel.
Make sure that you use the suitable partition panel plug for the
capillary.
TIP For capillaries with small outer diameter, the plug with slit secures
the capillary properly. For capillaries that do not fit in the plug with slit,
such as an insulated inlet capillary, use the rotating plug.
3. Connect the inlet capillary to the flow cell inlet port (IN).
1. See
Flow Connections to the Flow Cell
Figure 24: Connecting the detector waste line
1. Connect the waste line to the flow cell outlet (OUT) (1).
2. Route the waste line through the top recess (2) in the partition
panel.
3. Route the waste line through the tubing guides of the system
modules below the detector to the Vanquish system base (3).
4. On the system base, route the detector waste line through the
dedicated detector waste outlet to waste and connect the detector
waste line to the waste container as described in the Vanquish
System Operating Manual.
TIP The waste line should go straight to the system base and to waste.
Make sure that the line is positioned straight in the tubing guides.