Installation
5.3.3

Condensation

Page 50
WARNING—Electric Shock or Damage to the Device
Never use a power cord other than the power cords provided by
Thermo Fisher Scientific for the device.
Only use a power cord that is designed for the country in which you
use the device.
Do not use extension cords.
Never plug the power cord to a power socket that is shared with
other equipment (for example, multiple sockets).
Operate the device only from a power outlet that has a protective
ground connection.
In case of emergency, it must be possible to reach the power cord
easily at any time to disconnect the device from the power line.
WARNING—Electric Shock or Damage to a Product
Misuse of the power cords could cause personal injury or damage the
instrument. Use the power cords provided by Thermo Fisher Scientific
only for the purpose for which they are intended. Do not use them for
any other purpose, for example, for connecting other instruments.
NOTICE—Condensation in the device can damage the electronics and
optics.
When using, shipping, or storing the device, avoid or minimize
conditions that can lead to a build-up of condensation in the device.
For example, avoid significant or fast changes in environmental
conditions.
If you suspect that condensation is present, allow the device to
warm up to room temperature. This may take several hours. Wait
until the condensation is gone completely before connecting the
device to the power line.
Diode Array Detectors (VH-D10)
Operating Manual