7
Maintenance and Service
7.7
Flow Cell
To store or ship the flow cell, follow these steps:
To exchange the flow cell, follow these steps:
7.7.1
Preparing the Flow Cell for Storage
When
Parts required
Preparations
Page 126
This section describes how to flush and exchange flow cells.
1. Flush the flow cell (see
(} page 126)).
2. Remove the flow cell from the device and place it in its packaging
(see
Removing the Flow Cell
1. Remove the flow cell from the device (see
(} page 129)).
2. Install the new flow cell to the device (see
(} page 130)).
Before the flow cell is stored in its packaging or in the device if one of
the following applies:
Flushing the flow cell with highly pure isopropanol through the
system is not possible
It is shipped
Flushing and injection kit for flow cells, including a syringe and a
Viper adapter
Highly pure solvent that is miscible with isopropanol, such as LC/MS-
grade acetonitrile or methanol
Highly pure isopropanol, such as LC/MS-grade isopropanol
Plugs for the flow cell inlet and outlet (supplied with the flow cell)
1. Remove buffer salts by flushing the flow cell with a mixture of 50%
solvent and 50% water.
2. Flush the flow cell with a highly pure solvent that is miscible with
isopropanol, such as LC/MS-grade acetonitrile or methanol.
Preparing the Flow Cell for Storage
(} page 129)).
Removing the Flow Cell
Installing the Flow Cell
Diode Array Detectors (VH-D10)
Operating Manual