Allowed Concentrations; Further Information - Thermo Scientific Vanquish VH-D10 Operating Manual

Diode array detectors
Table of Contents

Advertisement

2
Safety
2.4.3

Allowed Concentrations

2.4.4

Further Information

Page 28
Allowed concentrations (standard system configuration):
System
Chloride
(Standard
Configuration)
Vanquish Core
0.1 mol/L
or less
Vanquish Horizon
1 mol/L or
less
Vanquish Flex
For details about the materials that are used in the analytical flow
path of the device, see the Specifications chapter in this manual. For
information about the materials that are used in the flow path of the
other modules in the Vanquish system, refer to the Specifications
chapter in the Operating Manual for the modules.
Observe the general guidelines and recommendations on the use of
solvents and additives in the chromatography system. Refer to Use
of Solvents and Additives in the Vanquish System Operating Manual.
Refer also to the Operating Manuals for all modules in the Vanquish
system. They may provide additional guidelines and information.
NOTICE
If the system configuration includes a non-standard detector, for
example, a charged aerosol detector or refractive index detector, refer
to the Operating Manual for the detector for specific recommendations
regarding solvents and additives.
Buffer
Remarks
1 mol/L or
High chloride concentration: The
less
application time should be as
short as possible. Flush the
-
system thoroughly after these
applications.
Diode Array Detectors (VH-D10)
Operating Manual