7.9

Replacing the Waste Line

When
Parts required
Tools required
Preparations
Follow these steps
Diode Array Detectors (VH-D10)
Operating Manual
NOTICE
Backpressures that exceed the specified maximum pressure limit of the
flow cell can destroy the flow cell. Observe the following:
Use only the waste line connection that is provided for your
detector.
Connect the waste line to the flow cell only as described in the
manual.
Do not discharge waste from the flow cell through the open leakage
drain system of the Vanquish system.
Never expose the flow cell to excessive backpressure.
Avoid clogging of the flow cell or waste line.
When connecting a component in the flow path after the flow cell,
observe the specified backpressure for the flow cell.
Blockage or leakage of one or more waste line components
Detector waste line
For instructions on connecting the waste line, follow the steps in this
section.
The waste line is routed through the system base and connected to the
waste. For instructions, refer to the Vanquish System Operating Manual.
Tubing cutter (optional)
1. Close the shutter in the light path of the flow cell.
2. Stop the pump flow to the flow cell.
1. When removing the waste line from the tubing guides, be careful
not to pull on other tubing in the guides.
2. Disconnect the waste line from the flow cell outlet.
7
Maintenance and Service
Page 145