Diode array detectors
Table of Contents

Advertisement

5
Installation
Page 58
2. Check the position of the rotating locks on the front of the flow cell.
If the rotating locks are not in a horizontal position, turn them
counterclockwise to a horizontal position. To insert the flow cell, the
rotating locks must always be in a horizontal position and thus
opened.
3. Carefully insert the flow cell into the flow cell opening in the device.
The flow cell must sit completely in the opening.
Figure 13: Inserting the flow cell with opened rotating locks
4. Turn the rotating locks simultaneously clockwise until they are in a
vertical position. The flow cell is locked in place, when the rotating
locks arrest.
Figure 14: Closing the rotating locks
5. Check that the flow cell sits correctly in the flow cell opening. The
front of the flow cell should lie flush with the device front panel.
Diode Array Detectors (VH-D10)