Tools required
Preparations
Follow these steps
Diode Array Detectors (VH-D10)
Operating Manual
Screwdriver, Torx T20
1. Unpack the replacement device. Remove the slide-in module from
its enclosure as described in
(} page 156).
2. Verify that the device enclosure into which the slide-in module shall
be installed is clean. If required, clean the inner and outer surfaces
of the enclosure. See
(} page 117).
3. When installing the slide-in module to an enclosure in the system
stack, check that the enclosure is placed correctly in the stack.
1. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to insert the slide-in module properly into the
enclosure in the next step.
2. On the slide-in module, check that you can move the captive screws
back and forth with your fingers. If you cannot, screw them in until
you can.
3. Insert the slide-in module in the enclosure. The following steps
require a team effort:
a) Take the slide-in module on both sides from below.
b) Lift the slide-in module to the height of the enclosure.
c) Place the slide-in module in the enclosure so that the module
sits in the enclosure by approximately 25 cm.
d) Push the slide-in module onto the rails and into the enclosure
until the slide-in module sits completely in the enclosure.
Figure 47: Inserting the slide-in module
7
Maintenance and Service
Removing the Slide-In Module
Cleaning or Decontaminating the Device