Messages - Thermo Scientific Vanquish VH-D10 Operating Manual

Diode array detectors
Table of Contents

Advertisement

8
Troubleshooting
8.2

Messages

Message and Code
Unexpected module behavior. Code
xx
Module malfunction detected. Code
xx
Code 22 X not detected
Code 33 Leak detected – eliminate
within approx. xx seconds.
Code 34 Leak detected.
Page 164
The table lists the most frequently observed messages for the device
and provides troubleshooting assistance.
Each message consists of a code number and a text. The code number is
the unique identifier for the problem while the wording may change.
Note the following:
To facilitate finding a message, the table lists the messages sorted by
code.
If you cannot find the code you are looking for, check the message
text. The two messages "Unexpected module behavior" and
"Module malfunction detected" can be assigned to different codes.
See the beginning of the table for more information.
TIP If you are unable to resolve the problem following the instructions in
this manual, or if you encounter a message not listed in the table, write
down the code and wording of the message and contact us. For details,
see the Contacting Us section at the beginning of this manual.
Description and Remedial Action
xx = Two-digit to four-digit code number.
When the message appears, write down the message code and turn
off the module. Wait for 5 seconds and turn on the module again.
TIP If the message appears with code 103, additional remedial actions
are available. See further down this table.
If the message appears again, contact Technical Support.
xx = two-digit to four-digit code number
When the message appears, write down the message code. Turn off
the module and contact Technical Support.
If the lamp is not detected, check whether the lamp is installed
properly. Make sure that you use a lamp that is provided or
recommended for your detector. If required, replace the lamp (see
Replacing the Lamp
contact Technical Support.
xx = the number of seconds within the leak must be resolved
Find and eliminate the source for the leakage (see
Leaks
(} page 170)).
Find and eliminate the source for the leakage (see
Leaks
(} page 170)).
(} page 122)). For all other undetected devices,
Diode Array Detectors (VH-D10)
Resolving Liquid
Resolving Liquid
Operating Manual