Thermo Scientific Vanquish VH-D10 Operating Manual page 144

Diode array detectors
Table of Contents

Advertisement

7
Maintenance and Service
Page 144
4. Proceed as required:
Situation
If a flow cell is to be
installed afterward
If no flow cell is to be
installed afterward
NOTICE
The flow cell opening on the device is sensitive to dust and debris.
If no flow cell is installed to the device, close the flow cell opening with
the cover for the flow cell opening.
Steps
Install the flow cell to the flow cell opening on the
device (see
Installing the Flow Cell
Install the cover to the flow cell opening on the device:
1. Check the position of the rotating locks on the cover.
The rotating locks must be in a horizontal, open
position. If required, turn the rotating locks
counterclockwise to a horizontal position.
2. Position the cover onto the flow cell opening. Make
sure that the cover is in the correct orientation.
3. Turn the rotating locks clockwise to a vertical
position to close the rotating locks and thus, to cover
the flow cell opening.
Diode Array Detectors (VH-D10)
(} page 130)).