5.8.5
Calculating the Pressure at the Inlet Port of the Flow Cell
Diode Array Detectors (VH-D10)
Operating Manual
6. Calculate the difference between the two measured pressure values:
p5 = p4 – p1.
p5: Backpressure caused by the flow cell
p4: System backpressure with flow cell but without additional
module
p1: Vanquish system backpressure (without flow cell and additional
module)
This procedure describes how to determine the pressure at the flow cell
inlet port (p6) for the intended configuration.
Figure 29: Calculating the pressure at the inlet port of the flow cell in
your used system configuration
No.
Description
1
Detector
2
Flow cell
2a
Outlet port of the flow cell
2b
Inlet port of the flow cell
3
Detector inlet capillary or
overpressure relief valve with
detector inlet capillary
4
Column
5
•
Installation
No.
Description
5
Waste container
6
Outlet port of the second
detector
7
Inlet port of the second
detector
8
Second detector
9
Waste line
10
Transfer capillary
Page 79