Status Indicators - Thermo Scientific Vanquish VH-D10 Operating Manual

Diode array detectors
Table of Contents

Advertisement

MUTE ALARM
UV
When the Device is Connected in the Chromatography Data System
6.3.2

Status Indicators

LED Bar
Diode Array Detectors (VH-D10)
Operating Manual
Beeping alerts you when the device detects a problem, for example,
leakage. To turn off the beep for the current alarm, press this button.
Eliminate the source for the alarm within 10 minutes. Otherwise,
beeping starts again. If the device detects a different problem, beeping
alerts you again immediately.
The UV button allows you to turn the UV lamp on and off directly from
the detector. To turn on the UV lamp, press the UV button.
The LED next to the button indicates the lamp status:
LED
Description
Off (dark)
The UV lamp is turned off.
Green, flashing
The UV lamp ignites.
Green
The UV lamp is turned on.
The button functionality is as follows when the device is connected in
the chromatography data system:
No injection or sequence or baseline monitoring is running:
All functions are available from the keypad.
An injection or sequence or baseline monitoring is running:
The MUTE ALARM function remains available from the keypad,
allowing you to turn off the beep for the current alarm.
The status LED bar on the front side of the device and the STATUS LED
on the inside keypad provide information about the device status.
The LED bar colors provide the following information:
LED Bar
Off (dark)
Dimmed
Yellow, flashing slowly
Description
The power to the device is turned off.
The doors of the device are open.
The power to the device is turned on, but the device
is not connected in the chromatography data
system.
6
Operation
Page 87