CAUTION—Spraying Solvent
Solvents can spray when under high pressure.
Stop the pump flow prior to opening the flow path.
Wait until the system pressure is down to zero.
When opening the flow path, wear appropriate protective
equipment.
CAUTION—Hot Surfaces
Surfaces inside the device may become hot during operation. Touching
hot parts might cause burns.
Allow hot surfaces to cool down before starting replacement or
maintenance procedures.
CAUTION—Hydrostatic Pressure
Solvent may spill when you open the flow path. This is due to hydrostatic
pressure in the system when the solvent reservoirs are located above
the pump outlet. Before you loosen a connection in the flow path:
Turn off the pump flow and wait until the system pressure is down
to zero.
Unscrew the caps of the solvent reservoirs and remove the solvent
lines together with the caps from the reservoirs.
Empty the solvent lines. Refer to the Operating Manual for the
pump.
Retighten the reservoir caps.
CAUTION—Electric Shock or Damage to the Device
After the power to the device is turned off, the device is still energized
as long as the power cord is connected. Repair work on the device while
the device is connected to power could lead to personal injury.
Always unplug the power cord before starting repair work inside the
device.
If you were instructed to remove any housing covers or panels, do
not connect the power cord to the device while the cover or panels
are removed.