Transporting The Device To A New Location - Thermo Scientific Vanquish VH-D10 Operating Manual

Diode array detectors
Table of Contents

Advertisement

7
Maintenance and Service

7.13.2 Transporting the Device to a New Location

Preparations
Follow these steps
Page 154
4. Remove the flow cell from the device and store it in its packaging.
Ensure that you install the flow cell cover to the flow cell opening.
Flow cells must be shipped in their original packaging.
NOTICE—Sensitive Flow Cells
Transporting the detector with a flow cell installed can lead to
destruction of the flow cell.
Therefore, remove the flow cell before transporting the detector.
NOTICE
The flow cell opening on the device is sensitive to dust and debris.
If no flow cell is installed to the device, close the flow cell opening with
the cover for the flow cell opening.
5. Remove the device or slide-in module from the system stack as
required:
Install the carrying handles and remove the device from the
Vanquish system. Follow the instructions on dismounting the system
stack in the Transporting or Shipping the System section of the
Vanquish System Operating Manual.
‑or‑
Remove the slide-in module from the device enclosure in the system
stack (see
Removing the Slide-In Module
Prepare the device for transport. See
(} page 153).
1. Observe the notes for handling and lifting the device safely.
2. Transport the device to the new location.
3. Install and set up the device in the system stack. Follow the
instructions on mounting the system stack in the Vanquish System
Operating Manual.
(} page 156)).
Preparing the Device for Transport
Diode Array Detectors (VH-D10)
Operating Manual