Thermo Scientific Vanquish VH-D10 Operating Manual page 67

Diode array detectors
Table of Contents

Advertisement

Parts required
Tools required
Preparations
Diode Array Detectors (VH-D10)
Operating Manual
Inlet capillary
Detector waste line
For instructions on connecting the waste line, follow the steps in
Connecting the Detector Waste Line
Plugs for the partition panel from the detector ship kit (if not pre-
installed), depending on the capillary to be installed:
For capillaries with small outer diameter, such as uninsulated
¨
capillaries, use the plug with slit (no. 1 in the image below).
For capillaries that do not fit in the plug with slit, including
¨
insulated capillaries, use the rotating plug (no. 2 in the image
below).
Figure 22: Plugs available for the partition panel
For the detector waste line: Tubing cutter (optional)
1. Flush the system modules and capillaries upstream of the device to
waste before you connect the flow cell to the system flow path.
Refer to the Vanquish System Operating Manual.
2. Remove the plugs from the flow cell inlet and outlet.
TIP Store the plugs of the flow cell, for example in the flow cell
packaging, to have them easily available when storing or shipping the
flow cell.
3. Install the partition panel plugs (if not present yet). See
Partition Panel Plugs
(} page 69).
(} page 62).
5
Installation
Installing the
Page 67