Probe Lever Replacement Procedure

8-2-5-1
Tools
Phillips screwdriver of appropriate size
FRU Part #
Refer to
8-2-5-2
Time Required
5 min
8-2-5-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in
and place it on a previously prepared surface with the display facing down.
8-2-5-4
Probe Lever Removal Procedure
1) Remove back cover.
2) Remove C-PSB.
3) Turnover the removed C-PSB unit, and identify the 2 screws holding the requested probe lever cap.
Note:
Each lever is held by 2 screws
Figure 8-81 Turn-over the Removed C-PSB Unit; Identify the Probe-lever Screws