LCD Flex Cable Replacement Procedure
8-3-8-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Flex Cable replacement
procedure.
8-3-8-2
Time Required
15 minutes
8-3-8-3
Preparation
Shut down the Vivid i n ultrasound unit, as described in
8-3-8-4
LCD Flex Cable Removal Procedure
1) Disconnect the AC cable and battery.
2) Remove the Control Panel and Keyboard as described in the
Removal Procedure" on page
3) Remove the LCD Display Frame as described in the
page
4) Remove the LCD Display as described in the
8-50
5
EVISION
8-22.
Section 8-3 - Internal Component Replacement Procedures
Chapter
8-6.
"LCD Display Frame Removal Procedure" on
"LCD Display Removal Procedure" on page
V
S
IVID I N
ERVICE
3-
System
Setup.
"Control Panel and Keyboard
M
ANUAL
8-45.