Section 8-11
Replacement of the Probe Holder
Replacement of the probe holder depends on the used trolley (GW).
NOTICE
• GW130, not to be replaced by the user
• GW140, see:
8-11-1
Replacement of the Probe Holder (at GW140)
8-11-1-1
Manpower
One person, 1 min.
8-11-1-2
Probe Holder - Replacement Procedure
1.) Check systems trolley version (refer to
2.) Remove the probe holder from the bottom housing of the User Interface.
3.) Place the new probe holder on the bottom housing.
8-22
DRAFT (A
Section 8-11-1 on page 8-22
Section 8-11 - Replacement of the Probe Holder
28, 2008)
PRIL
Table 9-3 on page
9-5). It has to be a GW140.