Halo Handle Replacement Procedure - GE Venue R1 Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5771498-100, R
IRECTION
8-6-3

Halo Handle Replacement Procedure

8-6-3-1
Tools
LOCTITE® 243™
Allen key 3 mm.
FRU Part #
Table 9-6
Refer to
8-6-3-2
Time Required
30 min
8-6-3-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-3-4
Halo Handle Removal Procedure
1.) Remove all accessories.
2.) Remove cockpit (monitor).
3.) Remove the arm.
4.) Remove the Riser Thermal Cover.
5.) Remove the Riser Cover.
6
EVISION
.
on page 9-5
Accessories - Replacement Procedures
Cockpit (Monitor) Removal Procedure
Arm Removal Procedure
Riser Thermal Cover Removal Procedure
Riser Cover Removal Procedure
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL
8-83