Up/Down Handle Replacement Procedure - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-6-6

Up/Down Handle Replacement Procedure

NOTE:
This section describes two removal procedures (for both the old [plastic] and new-type [metal] handles)
and one installation procedure (for the new-type handle), as follows:
-
-
-
8-6-6-1
Tools
Use the appropriate the Phillips-type screw drivers, as indicated in the up/down handle removal
procedure.
8-6-6-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
8-6-6-3
Up/Down Handle Removal - Procedure 1
NOTE:
This describes the procedure for removing the old-type (plastic) Up/Down Handle.
1) Raise the control console to its maximum height. Remove the bottom keyboard cover, as described
in
8-2-8-3 "Bottom Keyboard Cover Removal Procedure" on page
2) Loosen and remove the nut on the M6 screw (see
handle in position.
04
EVISION
Removal of the old-type (plastic) Up/Down Handle - see
Procedure
1, below.
Removal of the new-type (metal) Up/Down Handle with support - see
8-6-6-4 "Up/Down Handle Removal Procedure 2" on page
Installation of the Up/Down Handle with support - see
Procedure" on page
8-102.
Figure 8-82 Up/Down Handle Securing Mechanism
Chapter 8 Replacement Procedures
V
3N P
/E
IVID
RO
XPERT
Up/Down Handle Removal -
8-101.
8-6-6-5 "Up/Down Handle Replacement
Chapter 3 -
8-14.
Figure 8-82
below) that secures the up/down
M6 Nut
S
M
ERVICE
ANUAL
Installation.
Up/down Handle
M6 Screw
8-99