Ecg Module Replacement Procedure; Tools; Preparation; Bep2 - Ecg Module Removal Procedure - GE Vivid 4 BT03 RFI Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
8-5-13

ECG Module Replacement Procedure

NOTE:
In this section, the replacement procedure for an ECG Module for BEP1 and BEP2 systems are
provided in separate sub-sections.
8-5-13-1

Tools

Use the appropriate flat and Phillips-type screw drivers and a wire cutter.
8-5-13-2

Preparation

Shut down the Vivid™ 4 ultrasound unit, as described in
When ordered as a spare part (P/N: 22556476), the ECG kit will include the following items:
P/N
2334537
2552982
2274035
2274036
280A0306
280P0003
2298089
22771878
2334537INS
8-5-13-3

BEP2 - ECG Module Removal Procedure

NOTE:
BEP2 refers to a Pentium 4-based BEP (newer type).
1) Disconnect the mains AC power cable for the mains power wall outlet.
2) Remove the Vivid™ 4 left and right side covers - refer to the
page 8 - 4.
3.) Remove the Vivid™ 4 rear cover - refer to the
4) Disconnect all cables from the rear of the BEP.
5) Remove the BEP from the machine as described
page 8 - 48.
6) Remove both BEP side covers as described in the
8 - 54.
7) Remove the BEP front cover as described in the
8 - 54.
A left view of the exposed BEP showing the location of the ECG module is seen in
page 8-71.
8-70
3
EVISION
Patient I/O Assy. Box
Internal Foot Switch Cable
Slider Left
Slider Right
P.H.P Screw M3XL6
Spring Washer Int. M3
ECG Adaptor to COM2 Cable
Purpose Gaskets 10 cm strip
ECG Module Replacement Instructions
Section 8-5 - Back End Parts Replacement
Chapter 3 -
Description
Side Covers Removal Procedure
Rear Cover Removal Procedure
Back End Processor Replacement Procedure
BEP2 - Side Cover Removal Procedure
BEP2 - Front Cover Removal Procedure
V
™ 4 S
IVID
ERVICE
Installation.
Quantity
1
1
1
1
4
4
1
1
1
on page 8 - 7.
Figure 8-53
M
ANUAL
on
on
on page
on page
on

Advertisement

Table of Contents
loading

Table of Contents