Ecg Module Replacement Procedure - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-5-12

ECG Module Replacement Procedure

NOTE:
In this section, the replacement procedure for an ECG Module for BEP1 and BEP2 systems are
provided in separate sub-sections.
8-5-12-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter.
8-5-12-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
8-5-12-3
BEP1 - ECG Module Removal Procedure
NOTE:
BEP1 refers to a Pentium III-based BEP (earlier type).
1) Disconnect the mains AC power cable for the mains power wall outlet.
2) Remove the Vivid 3N left and right side covers - refer to the
Procedure" on page
3.) Remove the Vivid 3N rear cover - refer to the
7.
4) Disconnect all cables from the rear of the BEP.
5) Remove the BEP from the machine as described
Procedure" on page
6) Remove both BEP side covers as described in
Procedure" on page
7) Remove the BEP front cover as described in
page
8-80
04
EVISION
8-4.
8-57.
8-62.
8-63.
Section 8-5 - Back End Parts Replacement
V
3N P
IVID
Chapter 3 -
8-2-2-3 "Side Covers Removal
8-2-4-3 "Rear Cover Removal Procedure" on page 8-
8-5-1 "Back End Processor Replacement
8-5-2-3 "BEP1 - Left and Right Side Cover Removal
8-5-2-5 "BEP1 - Front Cover Removal Procedure" on
/E
S
M
RO
XPERT
ERVICE
ANUAL
Installation.