GE M
S
EDICAL
YSTEMS
D
2392751-100, R
IRECTION
8-5-5
Plug & Scan Battery Replacement Procedure
NOTE:
IMPORTANT - If the Plug & Scan battery is replaced, ensure that you properly transport and dispose
of the old battery according to the procedures established by your facility, company, or regional
regulatory agency.
8-5-5-1
Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrenches,
as indicated in the Plug & Scan battery replacement procedure.
8-5-5-2
Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in
the system when a Plug & Scan is present. Ensure that a full shutdown of the system is performed,
otherwise voltage might be present that can cause injury and/or damage.
8-5-5-3
Plug & Scan Battery Removal Procedure
1) Remove the left cover, as described in the
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP cover as described
The Plug & Scan battery and its cover are located to the right of the BEP.
4) Remove the two screws that secure the battery bracket in position (see
the bracket.
5) Disconnect the battery cable from the Plug & Scan board.
8-58
3
EVISION
Battery Bracket
Figure 8-48 Plug & Scan Battery Bracket and Screws
Section 8-5 - Back End Parts Replacement
Chapter 3 -
Side Covers Removal Procedure
Back End Processor Replacement Procedure
V
™ 4 S
IVID
ERVICE
Installation. Voltage may exist in
on page 8 - 4.
on page 8 - 48.
Figure 8-48
below); remove
M
ANUAL