FRU Replacement Procedures

Disassembly Guidelines

Battery Replacement

5-6
Maintenance: FRU Replacement Procedures
Prior to performing any disassembly procedures, perform the following:
If possible, process any ECGs remaining in storage.
If possible, print out set-up for future reference.
Disconnect the unit from the AC wall outlet and remove the power cord from
the unit.
Remove the battery as described in
If the MAC 3500 system is mounted to the trolley, remove it from the trolley as
described in
"Remove MAC 3500 System From
Remove the writer paper.
Take strict precautions against electrostatic discharge damage.
1.
Press the writer button to open the unit.
2.
Slide the battery release button in the direction of the arrow and lift the battery
out.
MAC™ 3500 Resting ECG Analysis System
2021337-036
"Battery
Replacement".
Trolley".
writer button
12A
Revision L 