Bios Battery Replacement - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-3-4

BIOS Battery Replacement

NOTE:
Since the BIOS battery is not available as a spare part (due to shelf life), it should be obtained locally.
The battery model is CR2032.
8-3-4-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the BEP replacement procedures.
8-3-4-2
Time Required
15 minutes
8-3-4-3
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-4-4
BIOS Battery Removal Procedure
1) Remove the BEP, as described in the
2) Take out the BIOS battery located on the bottom of the BEP, as shown in
8-3-4-5
BIOS Battery Installation Procedure
NOTE:
Before proceeding, refer to the
1) Fit a new BIOS battery in position on the BEP - refer to
2) Re-install the BEP, as described in the
3.) Turn ON the system and boot-up.
4.) Adjust to the correct date and time settings - see
5.) Proceed to perform the following functionality test:
-
1
EVISION
Figure 8-54 BEP - Location of BIOS Battery
Preparation Procedure on page 8 -
Battery Parameters Test on page 7 - 102
Chapter 8 - Replacement Procedures
V
IVID I N AND
BEP Removal Procedure on page 8 -
Figure
BEP Installation Procedure on page 8 -
Date and Time Adjustments on page 3 -
V
N S
IVID Q
ERVICE
38.
Figure
8-54.
BIOS Battery
43.
8-54.
42.
M
ANUAL
71.
8-45