Battery Module Replacement Procedure

8-4-7-1
Tools
Appropriate Phillips screwdriver.
FRU Part #
Table 9-10
Refer to
8-4-7-2
Time Required
15 min
8-4-7-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-7-4
Battery Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: RS Probe Cover, Lower Front eTower cover, Right Side eTower
Cover, Left Side eTower Cover and MPB Door Cover:.
3.) Release the two captive screws securing each battery pack.
4.) Pull out the top and bottom battery packs.
6
EVISION
.
on page 9-10
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
MPB Door Cover Removal Procedure
Figure 8-30 Removing the Battery Pack
Chapter 8 - Replacement Procedures
V
™ S
ENUE
Power Shut Down
.
M
ERVICE
ANUAL
Battery Pack
8-43