Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual page 151

Table of Contents

Advertisement

Thermo Fisher Scientific
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal components.
1. Remove the temperature control box cover per the previously
mentioned procedure
2. It is important to note or mark the location of all 16 cables that
connect to the temperature control board, to facilitate re-connection.
Disconnect all of those cables from the board
3. Using the Phillips screwdriver, remove the 2 mounting screws located
on the center of the long edges of the board.
4. Note the orientation of the board. One at a time, carefully unsnap the
board from the 4 corner mounting posts and remove it.
5. Place the new board on top of the 4 corner mounting posts in the
previous orientation, then push down on the corners to lock the board
in place. Be careful not to pinch the wires of any of the various cables.
6. Using the Phillips screwdriver, re-install the 2 mounting screws.
7. Re-connect all of the cables to their proper locations on the
temperature control board.
8. Re-install the temperature control box cover per the above procedure
"Removing and Installing the
Preventive Maintenance and Servicing
"Removing and Installing the
Cover".
Omni FTIR Multi Gas CEMS Instruction Manual
Temperature Control Box Components
Cover".
(Figure
5–55).
5-67