IMACC Script Setting
CEMS_SampleMode
temperature and gas cell pressure. Move the cursor to the block and double
click. The subpanel named Modbus Register Setting pops up, as shown.
On the top, there are two buttons, named "setup" and "serial". Under the
"setup" configuration, in the box of "OpticalPath", FrontMain is chosen.
Move down the setting block to "Measurement Channel Mapping". There
are six rows of register information, including measurement data type, code
name, register address and a check box for writing to "con" file. Highlight
the target parameter such as "Sampleline" in the list. Double click it so that
its register setting subpanel pops up, as shown.
As seen, its data type is IEEE float. "Code" name can be edited to reflect
the parameter meaning. "Register" context is determined by Calogix setting
(0x0a1c, in this case). Go back to the "Modbus Register Setting" panel,
click the "serial" button on the top, and view the following setting panel, as
shown.
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
E-3