5. Using the 9/16-inch wrench and the 5/8-inch wrench, loosen the brass
nut that attaches the copper tubing to the inlet of the check valve and
disconnect that tubing
6. Using the 9/16-inch wrench and the 1/2-inch wrench, loosen the brass
nut that attaches the copper tubing to the union fitting (on the
adjacent side of the heated block from the check valve) and disconnect
that tubing
(Figure
7. Turn the heated block assembly back around to access the front of the
assembly.
8. Using the 9/64-inch hex wrench, remove the 6 mounting screws and
remove the top valve clamp block (the one that has the tubing pass
through it.). (See
Figure
9. Using the 9/64-inch hex wrench, remove the 6 mounting screws and
remove the bottom valve clamp block
10. Lift the sample solenoid valve and plumbing assembly out of the heated
block
(Figure
5–61). Note the orientation of the tubing and fittings to
facilitate reconnection.
Figure 5–61. Solenoid Valve Replacement