  • Page 1 Thermo Scientific Orion 960 Titrator Addendum...
  • Page 2 ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher. 1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir...
  • Page 3 Calibrate pH/ISE in the Chapter  Autosampler Mode The Thermo Scientific Orion 960 titrator with revision 5.6 or higher software allows the operator to automatically calibrate a pH or an ion selective electrode in the autosampler mode. A two point pH or ISE calibration is allowed in beaker position 1 and 2 of the autosampler.
  • Page 4 4. Before moving to the next method 960 will print the electrode channel, Eo, and slope. Note: After 3 minutes, the 960 will take a reading regardless of the stability criteria. A new slope and Eo will substitute the 940 slope and Eo. Thermo Scientific Orion 960 Titrator Addendum...
  • Page 5 (Calculation and Substitution into the Methods) The Thermo Scientific Orion 960 titrator with revision 5.6 or higher software offers automatic blank determination and substitution for techniques 1, 2, 3, 6, 7, 8, 9, and 10. For example, a first derivative technique with mmol as a unit triggers a blank determination routine.
  • Page 6: Save Method

    CHOOSE METHOD xx ? Press YES or NO If YES is chosen, select up to 5 methods If NO is chosen, no method will be selected for blank substitution SAVE METHOD? Press YES or NO to continue Thermo Scientific Orion 960 Titrator Addendum...
  • Page 7 Concentration Calculations and Substitution into the Methods) The Thermo Scientific Orion 960 titrator with revision 5.6 or higher software offers standardization of the titrant concentration followed by automatic substitution of the mean value of the titrant in the selected saved methods.
  • Page 8 STANDARDIZE METHODS FAILED will be printed before starting a new method. • A seven digit code is added to the above new messages so DataPLUS2 software may receive them like other variables. Thermo Scientific Orion 960 Titrator Addendum...
  • Page 9 1: PRINTER 2: PC DATA COLLECT? has replaced with: ORION 960 CONNECTED TO A PRINTER? • In the new accessory menu, if “1” is selected then the 960 printer port can be used with the printer. If “2” is selected then the printer port on the 960 can be used for data collection with the DataPLUS2 software.
  • Page 10: Warranty

    United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/contactwater. Warranty For the most current warranty information, visit www.thermo.com/water. Thermo Scientific Orion 960 Titrator Addendum...