Electrical Panel
Components
Thermo Fisher Scientific
2. Using the 3/8-inch nut driver, remove the 2 nuts that mount the
control module and base unit assembly to the temperature control box.
Note the orientation of the assembly.
3. Note or mark the position of each temperature control module to
ensure that they are re-installed in the correct locations.
4. To remove one of the temperature control modules, lift the assembly so
you can access the bottom of it. Squeeze the two mounting tabs of the
module inward while lifting the module off of the base unit to release
it, taking care not to disconnect the wiring. Repeat to remove the other
3 modules.
5. Pull the two connectors off of the front of the temperature control base
unit, noting their location to facilitate reconnection.
6. Slide the din rail out of the bottom of the base unit to remove it.
7. Slide the din rail into the slots on the bottom of the new base unit until
it is centered.
8. Re-install each temperature control module into its appropriate slot in
the base unit by inserting the tabs into the slots and snapping them into
place.
9. Place the control module and base unit assembly onto the threaded
studs in the correct orientation. Thread the nuts on and tighten with
the 3/8-inch nut driver until snug.
10. Re-insert the 2 connectors in the front of the base unit in the
appropriate positions.
11. Re-install the temperature control box cover per the above procedure,
"Removing and Installing the
Use the following procedures to replace the electrical panel components
(Figure 5–56
and
Figure
Preventive Maintenance and Servicing
Cover".
5–57).
Omni FTIR Multi Gas CEMS Instruction Manual

Electrical Panel Components

5-71