Chapter 6
Troubleshooting
This chapter provides the following troubleshooting and service support
information for the Omni FTIR Multi Gas CEMS:
"Concentration
Errors" on page
"Boot Up Error/IMACC Software
"Temperature
Errors" on page
"Cell Pressure Errors/Sampling System
"Communication
Errors" on page
"Service
Locations" on page
The Technical Support Department at Thermo Fisher Scientific can also
be consulted in the event of problems. See
this chapter for contact information. In any correspondence with the
factory, please note both the model number and serial number of the
instrument.
Use the following check sheet to verify that parameters and settings are
within specification. This will assist in diagnosing any problems with the
system.
Check Sheet
Visual Inspection of Operating Parameters:
Pressure on system regulator on right side of rack:
System pressure = 75 PSI
Pressures on pneumatic panel:
CO
Free Air = 2.0 – 3.0 bar (30 – 44 PSI)
2
Pump Air = 1.0 – 3.0 bar (12 – 44 PSI)
Flows on pneumatic panel:
FTIR Purge Air = 1.0-1.5 LPM
6-3