Troubleshooting - Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual

Table of Contents

Advertisement

Troubleshooting

Service Locations
6-2
Omni FTIR Multi Gas CEMS Instruction Manual
Instrument Zero Air = 6 LPM when in instrument zero mode
Flowback = 3 LPM when in instrument zero mode
Probe Zero Cal Gas = 6 LPM when in probe zero mode
Probe Span Cal Gas = 6 LPM when in probe span mode
Probe Spike Cal Gas = ~0.5 LPM when in probe spike mode
Temperatures in IMACC software:
Heated sample line = 185 Deg. C
Heated pump/valve assembly = 180 Deg. C (after 30 minute
ramp-up)
FTIR Sample Cell = 180 Deg. C (after 2 hour ramp-up)
Internal Heated Lines (if installed) = 180 Deg. C
LED indicators on FTIR:
"Power" should be on continuously, if not check circuit breaker
and power supply
"Laser" should be on continuously, if not check laser assembly
"Source" should be on continuously, if not check source
"Scan" should be flashing during normal operation, if not check
if IMACC software is running
Readings in IMACC software:
FTIR Vmax > 0.9
FTIR Vmin < 0.0
If voltages are too small, check source, beamsplitter, detector,
cell windows and cell mirrors.
Cell Pressure = <40 mmHg less than ambient pressure during
sample mode
The following outline addresses possible instrument and alarm problems
and proposes checks and possible troubleshooting solutions.