About This Manual
WEEE Symbol
WEEE Symbol
iv
Omni FTIR Multi Gas CEMS Instruction Manual
Alert
Equipment Damage
The following symbol and description identify the WEEE marking used on
the instrument and in the associated documentation.
Symbol
Description
Marking of electrical and electronic equipment which applies to electrical and
electronic equipment falling under the Directive 2002/96/EC (WEEE) and the
equipment that has been put on the market after 13 August 2005.
Description
before removing the source from the analyzer. Handle the
source assembly only by the ceramic base.
The beamsplitters are extremely delicate. Hold them only
by the handle. Do not touch the mounting pads, and do
not touch or breathe on the substrate. Keep the exposure
to humid air to a minimum.
The following procedures must be followed in the correct
order. The ramp program must be saved and the program
mode must be set to 'On' before the controllers are set to
run. If the controllers are set to run with the ramp
program not enabled, heated line and sample cell
temperatures could rise too quickly and damage those
components.
Some internal components can be damaged by small
amounts of static electricity. A properly grounded
antistatic wrist strap must be worn while handling any
internal component.
If an antistatic wrist strap is not available, be sure to
touch the instrument chassis before touching any internal
components. When the PC is unplugged, the chassis is
not at earth ground.
When the temperature controller is unplugged, the
chassis is not at earth ground.