Diagnostics Menu
Access the Diagnostics menu to initiate a meter reset, test electrode stability, perform a meter self test or view meter
and module serial numbers and software revisions.
Main Menu
All meter settings are reset to factory defaults. Data log, calibration log and methods are deleted. Default password
Factory Reset
is 1111.
All settings for the selected channel are reset to factory defaults. Methods without a password are erased. Data log
User Reset
and calibration log are retained.
Tests the stability of a pH, ORP or ISE electrode by measuring the mV per minute drift and noise of the electrode.
Stability Test
Only channels with a BNC input can be selected.
Perform a keypad check, accuracy check (channels with BNC input only) and stirrer check (stirrer probe required). A
Self Test
self test is also performed automatically by the meter at each power up.
Displays the meter and module software revisions and serial numbers. The meter and module serial numbers and
About Meter
software revisions will be different.
Data Transfer
If measurements will be saved to the data log and then transferred to a computer or printer, make sure the data log
setting is on, the meter communication settings match the external device settings and the meter and external device
are properly interfaced.
The V
s
USB driver must be installed on the computer when interfacing the meter and computer via USB.
ersa
tar
™
™
™
Thermo Scientific
Orion
s
C
computer software facilitates the transfer of calibration and measurement data
tar
om
from the meter to a computer and can be downloaded for free at www.thermoscientific.com/OrionMeters.
thermoscientific.com/water
© 2014 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher
Scientific and its subsidiaries.
Water Analysis Instruments
North America
India
Toll Free: 1-800-225-1480
Tel: (91) 22-4157-8800
Tel: 1-978-232-6000
wai.asia@thermofisher.com
info.water@thermo.com
China
Netherlands
Tel: (86) 21-68654588
Tel: (31) 020-4936270
wai.asia@thermofisher.com
info.water.uk@thermo.com
Japan
Singapore
Tel: (81) 045-453-9175
Tel: (65) 6778-6876
wai.asia@thermofisher.com
wai.asia@thermofisher.com
IS-68X591201-B 0614
Settings
Australia
Tel: (613) 9757-4300
in Australia (1300) 735-295
InfoWaterAU@thermofisher.com
Thermo Scientific Orion
V
s
Benchtop Meter
ersa
tar
General Information
This literature provides basic instructions on operating the Thermo Scientific
more V
s
measurement module is installed. For comprehensive information on meter and module system setup,
ersa
tar
operation and advanced features, please refer to the V
or www.thermoscientific.com/water.
Preparation
Power Adapter
A universal power adapter with US, EU, UK and China plug plates is included
with the meter. Use of other power adapters can damage the meter and void the
warranty. The use of a surge protector or UPS is also recommended.
1. Select the appropriate plug plate for the power outlet.
2. Remove the clear plastic cover from the back of the power adapter.
3. Slide the plug plate into the groove on the back of the power adapter.
4. Connect the assembled power adapter to the outlet and meter input labeled
POWER.
™
™
™
Orion
V
s
meter when one or
ersa
tar
s
user manual available on the V
s
literature CD
ersa
tar
ersa
tar