Troubleshooting
Boot Up Error/IMACC Software Error
Boot Up
Error/IMACC
Software Error
Temperature
6-4
Omni FTIR Multi Gas CEMS Instruction Manual
Bad spectrometer detector or source
●
●
Concentration not updating in IMACC
●
●
H
●
●
System does not start automatically
●
●
Bad temperature readings – Either in 'Status' or in Modbus block in
IMACC script (See Appendix E,
Errors
"Last Temperature" reading in 'Status' on ImaccMonitor software
●
main screen reads 185.000000.
●
●
●
"Pump" sampling box temperature reading is low, but other
●
temperatures are OK.
●
ERROR_QUANT_TEMPRESSCON in ImaccMonitor
●
●
Check 'Last Vmax' and 'Last Vmin' in 'Status'. Nominally Last
Vmax >4.0 Volts, Last Vmin <-1.0 Volts. Readings of <±0.5
Volts indicate detector or source failure. Marginal readings may
require more detailed analysis by Thermo Fisher Scientific
Technical Support Department.
See 'Communications Errors/No Communications with FTIR'.
O concentrations <2000 ppm during zero checks
2
Make sure hydrator is not filled with water.
Verify the following programs are in the setup:
Omni.layout
ImaccCurrentLoop
ImaccModbus
ImaccModbusTCP
See
"System
Startup" in Chapter 2.
Check virtual com port number (See Appendix D,
Setup").
Check USB connection between PC and temperature
controller.
Stop and re-start script in ImaccMonitor.
Check thermostat inside pump box (See Chapter 5,
Maintenance and
Servicing".
Check temp control breaker and AC wiring connections.
"IMACC Script
Setting".)
"Calogix
"Preventive