Replacing The Pc; Figure 5-45. Heated Line Replacement - Thermo Scientific Omni FTIR Multi Gas CEMS Instruction Manual

Table of Contents

Advertisement

Replacing the PC

Thermo Fisher Scientific
Figure 5–45. Heated Line Replacement
8. Using the 9/16-inch wrench, loosen the nut and disconnect the heated
line(s) to be replaced from the bottom of the sample system heater
assembly, located at the back of the system rack. The 5/8-inch wrench
may be used to keep the bulkhead mounting nuts secure against the
heater box
(Figure
connection.
9. Install the new heated line(s) by following the previous steps in reverse
order. Attach insulation and tie wraps where appropriate.
Use the following procedure to replace the PC.
Equipment Required:
Phillips screwdriver
Flat blade screwdriver
PC
Preventive Maintenance and Servicing
5–45). Again, note their position to facilitate re-
Omni FTIR Multi Gas CEMS Instruction Manual
Replacing the PC
5-55